The design and manufacture of TYU-2000D DC Hysteresisgraph by Hunan Loudi Tianyu Electronic Co.,LTD completely meets the national standards GB3657-83 and the measurement verification standard JJG354-84、JJG407-86 and relevant IEC standard.
This kind of equipment has a big advantage that not only can obtain magnetic hysteresis loop and the parameter over the magnetic hysteresis loop by scanning the magnetic field but also can measure the characteristic of the quasi-static magnet in impact method. It made the magnetic measurement thoroughly get rid of dependent on ballistic galvanometer, so that it can avoid the non instantaneous error made by using ballistic galvanometer and it can save and develop the method of impulse measurement. The operation is very simple and the sensitivity can up to 4×10-8Wb, which already over the best of ballistic galvanometer.
Main Technical Specifications
|
Devicename |
TYU-2000D Type Soft Magnetic Direct Current Auto-testing Equipment |
|
Measuring Material |
Low coercivity soft magnetic materials, like Fe-Ni Alloy, Fe-Al alloys, Silicon Steel, Electrical Pure Iron and so on. High coercivity soft magnetic materials and weak magnetic materials like Magnetic Hysteresis Alloy and so on. |
|
Standard |
The National standard GB3657-83 and the measurement verification standard JJG35484、JJG407-86. |
|
Measurement Parameters |
Coercive force HC、Remanence Br、Saturation Magnetic Induction Intension Bs、Initial Permeability μi、Maximum Permeability μm、Magnetic Hysteresis Loss Pu、Relative Magnetic Permeability(weak magnetic materials);can it can draw magnetic hysteresis loop, permeability curve and initial magnetization curves. |
|
Sample Shape |
Open circuit state: rod and strip closed circuit status: loop、E、U shapes and so on. |
|
Measurement Method |
Impact method and scanning method there are classes: closed circuit and open circuit(±30V×15A,±40V×15A) |
|
Magnetization Device |
wire winding(closed circuit measurement) and solenoid(opened circuit measurement)、soft magnetic permeameter;it can offer 1500A/cm magnetic field. |
|
Systematic Error |
Magnetic Flux<±0.3%. Electric current<±0.3% |
The measuring error of the closed magnetic circuit samples
|
Measured Parameter |
Measurement Repeatability |
Uncertainty Degree |
|
Bs |
±0.3% |
1.0% |
|
Br |
±0.3% |
1.0% |
|
Hc |
±0.3% |
1.0% |
|
μi |
±1.5% |
4.0% |
|
μm |
±1.0% |
2.0% |
Open Circuit sample(adopt the samples of solenoid or permeameter) measuring the errors.
|
Measured Parameter |
Measurement Repeatability |
Uncertainty Degree |
|
Hc |
1.0% |
2.5% |
|
Bm |
0.5% |
1.5% |
|
Pu |
1.0% |
2.0% |
System Configuration
|
Components Name |
|
|
Computer |
The computer adopts the popular Standard configuration of the market. Customers can have some choices if there is other requirement. |
|
Printer |
Laser printer or color ink jet printer can be chose. |
|
A/D D/A |
RS232 or USB interface |
|
Excitation Power |
DC;30V;15A、over voltage、over current、short-circuit protection, high stability, superior to 0.05%; current output: 0.1~15000mA |
|
Integraph |
The measurement range is 0.1、2、5、10 mWb, automatic measurement range control. |
|
Standard Solenoid |
For opened circuit measurement(Fittings can be chose) |
|
Demagnetizer |
Adapt to soft magnetic material |
|
Measuring Software |
Adapt to work under WIN98/2000/XP |
|
Others |
The instruction of maintenance, Vulnerable component |
Attached Drawing of the Software

Above is the drawing of basic magnetization curve and the conductance curve measured in impact measuring method.
Above is the drawing of magnetic hysteresis loop measured in scan measuring method. |